Z-2063-2016: Liquid stable 2 part Homocysteine Reagent, for in vitro quantitative determination of total homocysteine

Recall record Z-2063-2016 for Medtest Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2063-2016
Company in FDA record
Medtest Holdings, Inc.
FDA record creation date
Jun 28, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Deterioration in performance of the Liquid Stable (LS) 2-Part HOMOCYSTEINE REAGENT resulting in under recovery. The manufacturer has indicated that the HOMOCYSTEINE REAGENT lot listed above shows deterioration in assay performance manifesting itself by under recovery of Homocysteine in Controls and Patient Samples. The cause of this deterioration has been attributed to a loss of reductant activity in Reagent 1. Testing performed indicated that the affected lot is at the limit of acceptable clinical performance. This observation was consistent across the reportable range of the assay. For this reason, do not continue to use affected reagent lot. Test data generated to date will be acceptable where validity control specifiCations have been met.

Source facts

Action
On 4/3/2015, Urgent Field Safety Notice notifications were sent to the affected customers via email with instructions for identifying and destroying the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification.
Initiated
2015-04-03
Root cause
Unknown/Undetermined by firm
Terminated
2016-07-05
Distribution
MI, MN, NJ, IL, NY. US Only.
Recall number
Z-2063-2016
Recall status
Terminated
CDRH record ID
146910
FDA record created
2016-06-28
Recall event number
74234

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗