Z-2063-2017: Healon GV, Part No. 10294701, 10294801, 10200014, 10201014, 10202014, 10203014

Recall record Z-2063-2017 for Abbott Medical Optics Inc. (AMO). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2063-2017
Company in FDA record
Abbott Medical Optics Inc. (AMO)
FDA record creation date
May 10, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Remote possibility that certain solutions in these lots may contain microscopic glass particles due to damage that occurred at the cylinder neck during the manufacturing process.

Source facts

Action
A recall letter was sent to customers on 4/1/17 to inform them that AMO has voluntarily initiated this Action because a remote possibility exists that certain Healon OVD solutions in the affected lots may contain microscopic glass particles due to damage that occurred at the cylinder neck during the manufacturing process. Use of OVD solutions with glass particles could potentially lead to intraocular injury. Customers are instructed to complete the Customer Reply Form and fax it to AMO Quality Assurance at 714-247-4510 or email to RegCompliOne@abbott.com within 3 business days of receipt of the letter. Customers with product complaints or adverse events are instructed to inform AMO by calling 877-266-4543. Customers that do report a complaint are instructed to provide the Healon OVD lot number and, if a patient was involved, the date of surgery, a description of the event and patient outcome.
Initiated
2017-04-01
Root cause
Process design
Terminated
2017-11-27
Distribution
US and worldwide: Austria Belgium Croatia Czech Republic Denmark Finland France Germany Great Britain Iceland Ireland Israel Italy Latvia Lebanon Lithuania Netherlands Norway Portugal Spain Sweden Switzerland Tunisia Turkey Russian Fed Guadeloupe Sri Lanka Australia China Hong Kong Indonesia Malaysia Singapore South Korea Taiwan Thailand Japan Chile Colombia Costa Rica Ecuador
Recall number
Z-2063-2017
Recall status
Terminated
CDRH record ID
154836
FDA record created
2017-05-10
Recall event number
77023

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗