Open original FDA record ↗Recall record
- FDA identifier
- Z-2064-2016
- Company in FDA record
- CareFusion 303, Inc.
- FDA record creation date
- Jun 28, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A patient weight can be populated incorrectly under certain conditions when using the RESTORE feature to restore infusions running on the Alaris LVP module model 8100 and the Alaris Syringe module model 8110.
Source facts
- Action
- An Urgent Medical Device Recall Letter was sent to customers on 5/31/16 to inform them that BD had identified an issue with the Alaris System PC unit model 8015 with software versions 9.17 and 9.19. Firm has received reports that a patient weight can be populated incorrectly under certain conditions when using the RESTORE feature to restore infusions running on the Alaris LVP module model 8100 and the Alaris Syringe module model 8110. The letter informs the customers of the potential risk and the required action for users. Customers are informed that BD will contact them within 60 days to initiate the scheduling process for the software upgrade installation.
- Initiated
- 2016-05-12
- Root cause
- Software design
- Terminated
- 2018-12-24
- Distribution
- Distributed US (nationwide) and in Canada.
- Recall number
- Z-2064-2016
- Recall status
- Terminated
- CDRH record ID
- 146476
- FDA record created
- 2016-06-28
- Recall event number
- 74142