Open original FDA record ↗Recall record
- FDA identifier
- Z-2065-2012
- Company in FDA record
- GE OEC Medical Systems, Inc
- FDA classification date
- Jul 24, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class I
Reason stated in the source
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The OEC 8800 Flexview was recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system lok-up, incorrect image display, incorrect dose area calculated air kerma in "film mod
Source facts
- Action
- Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE Healthcare sent two Urgent Recall Notice letters to the Facility Administrator and the Radiology Administrator. The letters identified the affected product, problem and actions to be taken. The letters advised consignees that GE Healthcare is actively working on solutions that will permanently resolve each of the issues. For questions or concerns contact GE Healthcare service team at 800-874-7378 option 8.
- Initiated
- 2006-11-20
- Root cause
- Device Design
- Terminated
- 2013-05-13
- Distribution
- Worldwide Distribution - US Nationwide
- Recall number
- Z-2065-2012
- Recall status
- Terminated
- CDRH record ID
- 110394
- FDA classified
- 2012-07-24
- Recall event number
- 62387