Z-2065-2026: LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 9957

Recall record Z-2065-2026 for Physio-Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2065-2026
Company in FDA record
Physio-Control, Inc.
FDA classification date
May 4, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.

Source facts

Action
On 04/13/2026, recall notices were sent to customers who were asked to do the following: 1) Return the business reply form by email to RSRecall@Stryker.com to confirm receipt of this notification. 2) Upon receipt of the completed business reply form, firm will contact you to arrange for corrective action on the device if service has not been performed. 3) Continue to follow the Operating Instructions and perform the daily checks and user tests. These tests can be found in the chapter "Maintaining the Equipment" subsection "General Maintenance and Testing". 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. Please ensure this letter is kept with the affected device until the correction has been completed. 5) Inform firm if any of the subject devices have been distributed to other organizations. If so, provide contact details, so the firm can inform the recipients. If you have any questions or concerns, please contact the firm's Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday - Friday or by email at medtechsup@stryker.com
Initiated
2026-04-13
Root cause
Process control
Distribution
US Nationwide distribution in the states of MI, WA, KY, IN, FL, NC, TX, LA, VA, NY, MA, CO, PA, OH, SC, OK, MN, WV, NJ, WI.
Recall number
Z-2065-2026
Recall status
Open, Classified
CDRH record ID
218656
FDA classified
2026-05-04
Recall event number
98418

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗