Skip to content

Public FDA record

Z-2066-2026: Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adap

Recall record Z-2066-2026 for Windstone Medical Packaging, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2066-2026
Company in FDA record
Windstone Medical Packaging, Inc.
FDA classification date
May 8, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.

Source facts

Action
Firm began notifying consignees via email on April 2, 2026. Notice was titled "URGENT: MEDICAL DEVICE RECALL." Consignees were instructed to check inventory and quarantine affected products. Firm instructed consignees to over-label affected products with a sticker which states that at the time the kit is opened, the affected Medline syringe should be identified and set aside, and that the affected syringe should be rendered unusable and disposed of. Firm will provide labels for any kits that need to be identified.
Initiated
2026-04-02
Root cause
Device Design
Distribution
US Nationwide distribution in the state of Arizona.
Recall number
Z-2066-2026
Recall status
Open, Classified
CDRH record ID
219675
FDA classified
2026-05-08
Recall event number
98741

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗