Open original FDA record ↗Recall record
- FDA identifier
- Z-2066-2026
- Company in FDA record
- Windstone Medical Packaging, Inc.
- FDA classification date
- May 8, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.
Source facts
- Action
- Firm began notifying consignees via email on April 2, 2026. Notice was titled "URGENT: MEDICAL DEVICE RECALL." Consignees were instructed to check inventory and quarantine affected products. Firm instructed consignees to over-label affected products with a sticker which states that at the time the kit is opened, the affected Medline syringe should be identified and set aside, and that the affected syringe should be rendered unusable and disposed of. Firm will provide labels for any kits that need to be identified.
- Initiated
- 2026-04-02
- Root cause
- Device Design
- Distribution
- US Nationwide distribution in the state of Arizona.
- Recall number
- Z-2066-2026
- Recall status
- Open, Classified
- CDRH record ID
- 219675
- FDA classified
- 2026-05-08
- Recall event number
- 98741