Reason stated in the source
Accessories to the GYC-1000 laser were missing Laser Aperture labels.
Source facts
- Action
- Nidek sent Recall Notifications to doctors on March 31, 2015, by Federal Express. Letters were specific to the accessory involved. NIDEK detected that the laser aperture label has not been affixed on the delivery unit of GYC-1000, Green Laser Photocoagulator System and the device does not meet FDA regulation 21 CFR Subchapter J. In order to comply with FDA regulation, Nidek requests that the customer affix the enclosed laser aperture label to the Dual Delivery 4DD-1 of GYC-1000, according to the instructions attached. Please see the enclosed Laser Aperture label, instruction for affixing to the device and the response card. Nidek requests that you complete and return the response card as soon as possible. For additional information, please call Nidek at 1-800-223-9044 or visit http://usa.nidek.com
- Initiated
- 2015-03-25
- Root cause
- Error in labeling
- Terminated
- 2015-07-14
- Distribution
- US Distribution to the states of : CA, IL, MN, MO, OR, IL, UT RI, MA, WA, NE, NY, NJ, TX and WI.