Z-2067-2019: Centurion Convenience Kits, containing BD Alaris Pump Infusion Sets. BD code 11522558 (Centurion code 110

Recall record Z-2067-2019 for Medline Industries Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2067-2019
Company in FDA record
Medline Industries Inc
FDA classification date
Aug 12, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

This recall is a sub recall to BD s recent Class I recall for the Alaris Pump Infusion Set due to an issue with the consistency of the inner lumen thickness of the silicone tubing segment which directly impacts amount of occlusion force necessary to fully seal the tube. This defect prevents the pump from complete occlusion and results in the potential for over-infusion of drug product into the patient.

Source facts

Action
On May 10, 2019, the firm notified its distributor, Medline, of the recall. The distributor was asked to identify all inventory within its possession, and to cease further distribution. The distributor was asked to notify any customers to whom the recalled product had been distributed, and to instruct customers to cease use immediately. On May 13, 2019, Medline notified its customers of the recall. Medline's instructions to customers were as follow: 1. Immediately check your stock for the specific kit number and the affected kit lot number in the attached list. Quarantine any kits with the listed lot numbers. If you dont have any of the lots listed above skip to step 3. 2. Affix the enclosed stickers on the affected kits and return to stock. If additional stickers are needed please call 866-359-1704. Please be sure NOT to use the recalled Infusion Set components, pull another approved product from sterile supply. 3. Please fill out and return the enclosed Verification Form via email to recalls@medline.com or fax 866-767-1290. Customers with questions may call 1-866-359-1704.
Initiated
2019-05-10
Root cause
Nonconforming Material/Component
Terminated
2021-08-18
Distribution
One distributor located in Illinois.
Recall number
Z-2067-2019
Recall status
Terminated
CDRH record ID
173181
FDA classified
2019-08-12
Recall event number
82951

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗