Z-2068-2020: DFINE Vertebral Assemblies: Hydraulic-Master Syringe Assembly, StabiliT Vertebral Augmentation System, Re

Recall record Z-2068-2020 for Merit Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2068-2020
Company in FDA record
Merit Medical Systems, Inc.
FDA record creation date
May 21, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Vertebral hydraulic assemblies has incorrect expiration date printed on the labels.The product's shelf-life is two (2) years, but the expiration date printed on the labels is five (5) years.The affected product will begin expiring in March 2020. Using product beyond the validated shelf-life may result in (1) product not functioning as intended and/or (2) a delay in procedure or the need for the procedure to be rescheduled.

Source facts

Action
On 02/13/2020, the firm notified affected customers using a written notice titled, "URGENT PRODUCT RECALL NOTICE". The letter indicated: (1) Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. (2) Ensure that applicable personnel within your organization are made aware of this field action. (3) If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. (4) Please fill out, scan and email the completed Customer Response Form to Customer Service at response email address indicated in the letter. All affected product shipped to you must be accounted for on the CRF. (5) Please immediately return all affected lots in your possession via Federal Express 2nd Day Air using the account number provided in the letter. Please include a copy of the Customer Response Form with the returning product, reference the assigned RMA number on the outside of the box (see Customer Response Form), and ship to ATTN: Receiving & Customer Service at the firm's address listed in the letter. If you have any questions concerning this communication, please don't hesitate to contact your Sales Representative or Customer Service at (801) 208-4381.
Initiated
2020-02-13
Root cause
Labeling Change Control
Terminated
2021-02-26
Distribution
Worldwide distribution: US Nationwide distribution including in the states of (TX, UT, PA, KY, TX, AR, CA, and PA) which included one (1) government/military consignee in KY; and the countries of (Germany, Switzerland, and New Zealand).
Recall number
Z-2068-2020
Recall status
Terminated
CDRH record ID
181059
FDA record created
2020-05-21
Recall event number
85485

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗