Skip to content

Public FDA record

Z-2068-2026: Mint Lesion; Software Versions: 3.4.0 up to 3.9.5.;

Recall record Z-2068-2026 for Mint Medical GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2068-2026
Company in FDA record
Mint Medical GmbH
FDA classification date
May 6, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If the connection between a mint Lesion workstation and the mint Lesion server is interrupted while performing a read, in rare cases some information may be lost or incorrectly linked to other patients.

Source facts

Action
On or about January 7, 2026 Urgent Field Safety Notice letters were emailed to customers. Customers were instructed to read the notice and check if any affected product is available for use. If that is the case, the malfunction may have occurred and may occur in your system. Please be aware that the malfunction may occur. Your IT department must ensure that no temporary network interruptions occur in your clinical network. Also, your IT department must ensure that the mint Lesion server is not restarted while users work with mint Lesion application (i.e., the server may be restartet, but only outside of working hours). Please verify correctness of the information of prior time-points. Be attentive to potential inconsistencies in lesion attributes when opening existing cases for follow up. Cross check lesion location and categorization with prior radiological reports and image data. If you suspect that the problem may have occurred on your mint Lesion system, please contact Mint Medical support via support@mint-medical.com or any other contact details for technical analysis and investigation. Mint Medical Support will analyze if the malfunction appeared at your mint Lesion system in the past. If you believe that this failure has occurred, please review the potentially affected radiological reports in your reporting application and take the necessary steps to correct them. Pass this notice on to all users of mint Lesion" within your organization. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action.
Initiated
2026-01-07
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Germany, Switzerland, Austria, Belgium, Italy, France, Romania, Netherlands, United Kingdom, Spain, Czech Republic, Ireland, Poland;
Recall number
Z-2068-2026
Recall status
Open, Classified
CDRH record ID
219696
FDA classified
2026-05-06
Recall event number
98753

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗