Open original FDA record ↗Recall record
- FDA identifier
- Z-2069-2017
- Company in FDA record
- Almore International Inc
- FDA classification date
- May 10, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Microfil Composite Instruments were distributed with a sterile symbol on their packaging label inserts, indicating that the products are sold as sterile devices. These products are not sterile and they are intended to be sterile prior to each use.
Source facts
- Action
- Almore International sent an urgent medical device recall letter via certified mail on April 20, 2017 to consignees informing them of the labeling error and that the device is not sterile upon receipt. On 5/8/17, the firm expanded the recall and will notify four additional consignees by 5/10/17. Almore will send new replacement label inserts to customers who still have affected device on hand upon request. Distributor should notify downstream customers and/or provides new label inserts to downstream customers as needed. Consignees should send response form to Almore International, Inc. using one of the following methods below: o Fax the completed form to 503-643-9748 or o Send the scanned form via email to info@almore.com or o Mail a copy of the completed form to: Microfil Recall, Almore International Inc. 10950 SW 5th Street, Suite 270 Beaverton, OR. 97005. If consignees have questions please contact Almore International, Inc. at 503-643-6633 Monday through Friday between 9am and 4pm PST.
- Initiated
- 2017-04-20
- Root cause
- Employee error
- Terminated
- 2017-08-22
- Distribution
- Worldwide distributed - US Nationwide in AZ, CA, CO, CT, FL, GA, IA, IL, IN, MA, MN, MO, NC, NJ, NV, NY, OH, PA, TN, TX, WA, and WI, and internationally in Canada, Netherlands, and South Korea.
- Recall number
- Z-2069-2017
- Recall status
- Terminated
- CDRH record ID
- 154721
- FDA classified
- 2017-05-10
- Recall event number
- 76963