Z-2069-2019: Versa - Kath Mini Kit, with plastic L.O.R Syringe 12g X 12in (30.5 cm), 5 units, Sterile ED, REF 181-211

Recall record Z-2069-2019 for Epimed International, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2069-2019
Company in FDA record
Epimed International, Inc.
FDA record creation date
Jul 24, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Mislabeled - The product's labeling contains an incorrect size for the catheter included in the kit. The kit contains a 21 gauge catheter, however the labeling states a 12 gauge catheter is included.

Source facts

Action
On June 19, 2019 , the firm mailed a lot specific, medical device market correction letter, via USPS to consignees who received the affected product(s). The letter asked customers to do the following: 1. Please examine your inventory and quarantine any product subject to the recall immediately. 2. Once product has been Quarantined, please complete and return the response form as soon as possible. 3. Please contact EPIMED sales administrative assistance, Bobbi Harrington at (518) 725-0209 ext. 1200 or bobbih@epimed.com, so that immediate replacement or credit for the affected product can be arranged. 4. If you have any questions, please contact Mr. Kris Knapp at 1-800-866-3342 ext. 1305.
Initiated
2019-06-18
Root cause
Labeling mix-ups
Distribution
Distribution in US states of CA, FL, OK, AR, WA, TX, ID, NY, South Africa, and Mexico
Recall number
Z-2069-2019
Recall status
Terminated
CDRH record ID
174228
FDA record created
2019-07-24
Recall event number
83247

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗