Open original FDA record ↗Recall record
- FDA identifier
- Z-2069-2020
- Company in FDA record
- Baylis Medical Company Inc.
- FDA classification date
- May 21, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a remote potential for the sterile packaging of lots of the transseptal guiding sheath to be compromised during shipping and handling. In the case that a packaging defect is not identified and the product is used in a patient, there may be a remote risk of infection.
Source facts
- Action
- The firm sent Urgent Medical Device Recall Notification letters to customers via UPS and/or by email on 07-Oct-2019. The notifications were followed through by a tracking number to facilitate confirmation of package delivery. Customers were informed that all Torflex Transseptal Guiding Sheaths sold separately or as part of the NRG Transseptal kit are affected by this action. Customers were instructed to examine their stock, determine whether they had any of the affected lot numbers in stock, and if so, to quarantine affected product. Customers were requested to complete the acknowledgement form and return it to Baylis within 3 business days by email at torflexfieldaction@baylismedical.com or by faxing the completed form to 906-602-5671 Attn: Quality Department. Customers were informed that once the completed Acknowledgement form was received by Baylis, they would be contacted by a Baylis Medical Representative within 5 business days to manage the disposition of any identified quarantined product. Customers were instructed to address specific questions regarding the recall to Mrs. Ellen Harfield, Director of Quality at 1-800-276-4416.
- Initiated
- 2019-10-07
- Root cause
- Package design/selection
- Terminated
- 2021-06-30
- Distribution
- Product distributed throughout the 50 states and in Washington D.C.
- Recall number
- Z-2069-2020
- Recall status
- Terminated
- CDRH record ID
- 180671
- FDA classified
- 2020-05-21
- Recall event number
- 85353