Open original FDA record ↗Recall record
- FDA identifier
- Z-2070-2014
- Company in FDA record
- Excelsior Medical Corp
- FDA classification date
- Jul 16, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential for a leak. A molding defect in the connector that joins the three leads to the main flow line may allow fluid to escape or air to enter the fluid path during transfer of fluids from one container to another.
Source facts
- Action
- Excelsior Medical issued "Urgent Medical Device Recall" notifications/"Customer Reply Forms" dated June 5, 2014 to affected customers via certified mail (return receipt request). The notice informed customers of the issue with the affected product; how to identify affected product; instructed the customers to discontinue use of tubing sets from the affected lots and return the attached form (Customer Reply Form) via fax to Excelsior to make arrangements for return of product. If product was further distributed by the customer the notification instructs them to identify the customer(s) that it was supplied to and notify them at once of the product recall. A phone number was provided in case customers had questions regarding the recall (1.800.487.4276 8:30 am - 5:00 pm EST) as well as an email address (recall12@excelsiormedical.com).
- Initiated
- 2014-06-05
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-08-19
- Distribution
- Worldwide Distribution - USA (nationwide) and Canada.
- Recall number
- Z-2070-2014
- Recall status
- Terminated
- CDRH record ID
- 128242
- FDA classified
- 2014-07-16
- Recall event number
- 68573