Z-2070-2020: NRG RF Transseptal Kit; Model No.s : NTK-710-8545, NTK-710-8555, NTK-711-8545, NTK-711-8555, NTK-711-8590

Recall record Z-2070-2020 for Baylis Medical Company Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2070-2020
Company in FDA record
Baylis Medical Company Inc.
FDA classification date
May 21, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a remote potential for the sterile packaging of the transseptal guiding sheath to be compromised during shipping and handling. In the case that a packaging defect is not identified and the product is used in a patient, there may be a remote risk of infection.

Source facts

Action
The firm sent Urgent Medical Device Recall Notification letters to customers via UPS and/or by email on 07-Oct-2019. The notifications were followed through by a tracking number to facilitate confirmation of package delivery. Customers were informed that all Torflex Transseptal Guiding Sheaths sold separately or as part of the NRG Transseptal kit are affected by this action. Customers were instructed to examine their stock, determine whether they had any of the affected lot numbers in stock, and if so, to quarantine affected product. Customers were requested to complete the acknowledgement form and return it to Baylis within 3 business days by email at torflexfieldaction@baylismedical.com or by faxing the completed form to 906-602-5671 Attn: Quality Department. Customers were informed that once the completed Acknowledgement form was received by Baylis, they would be contacted by a Baylis Medical Representative within 5 business days to manage the disposition of any identified quarantined product. Customers were instructed to address specific questions regarding the recall to Mrs. Ellen Harfield, Director of Quality at 1-800-276-4416.
Initiated
2019-10-07
Root cause
Package design/selection
Terminated
2021-06-30
Distribution
Product distributed throughout the 50 states and in Washington D.C.
Recall number
Z-2070-2020
Recall status
Terminated
CDRH record ID
181315
FDA classified
2020-05-21
Recall event number
85353

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗