- FDA identifier
- Z-2070-2025
- Company in FDA record
- HemoCue AB
- FDA classification date
- Jul 2, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Glucose microcuvettes experienced transit time outside of limits, so stability through the whole microcuvette lifetime can no longer be guaranteed. The affected products may not have been stored at the recommended labeled storage conditions which may have impacted the safety, quality, identity, potency, and purity of the product. This failure could increase the probability of an incorrect test result or a delayed test result.
Source facts
- Action
- On 5/14/2025 field notices were emailed to customers who were asked to do the following: 1) Ensure that the affected product at customer site has been destroyed according to local regulations. 2) Complete and return the Field Notice Verification Form. If you have any questions regarding this Field Notice, please send an e-mail to cuvettes@hemocue.com
- Initiated
- 2025-05-12
- Root cause
- Storage
- Distribution
- US Nationwide distribution in the state of AR.
- Recall number
- Z-2070-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214236
- FDA classified
- 2025-07-02
- Recall event number
- 97025