Skip to content

Public FDA record

Z-2072-2010: TomoTherapy Hi-Art System, ver. 4.0.0, 4.0.1, 4.0.2, ThomoTherapy Incorporated 1240 Deming Way, Madison

Recall record Z-2072-2010 for TomoTherapy Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2072-2010
Company in FDA record
TomoTherapy Incorporated
FDA classification date
Jul 22, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

An issue has been identified, as a result of input received from one TomoTherapy Hi-Art Treatment System customer. With a particular workflow, the Planning Station's Draft and/or final Plan Report can provide incorrect dose statistics.

Source facts

Action
Consignees were sent on 6/23/10 a TomoTherapy "Urgent Field Safety Notice Medical device Correction" letter dated June 22, 1010. The letter addressed the Issue, Product Affected, Recommended Action and Resolution. Prior to generating a draft or final Plan Report for a given plan, the plan must be closed and reopened at least once. This issue will be remedied by Hi-Art software version 4.0.3, scheduled for release in late 2010. Customers with questions were provided with a contact in the letter at the TomoTherapy Customer Interaction Center. The firm can be contacted at 608-824-2800.
Initiated
2010-06-23
Root cause
Software design
Terminated
2011-12-17
Distribution
CA, CO, IL, LA, MI, MO, NY, ND, OH, OK, PA,VA, WI, BELGIUM CANADA, KOREA, MALAYSIA, SWITZERLAND, ITALY, GREAT BRITAIN, FRANCE.
Recall number
Z-2072-2010
Recall status
Terminated
CDRH record ID
92555
FDA classified
2010-07-22
Recall event number
56056

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗