Open original FDA record ↗Recall record
- FDA identifier
- Z-2072-2010
- Company in FDA record
- TomoTherapy Incorporated
- FDA classification date
- Jul 22, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
An issue has been identified, as a result of input received from one TomoTherapy Hi-Art Treatment System customer. With a particular workflow, the Planning Station's Draft and/or final Plan Report can provide incorrect dose statistics.
Source facts
- Action
- Consignees were sent on 6/23/10 a TomoTherapy "Urgent Field Safety Notice Medical device Correction" letter dated June 22, 1010. The letter addressed the Issue, Product Affected, Recommended Action and Resolution. Prior to generating a draft or final Plan Report for a given plan, the plan must be closed and reopened at least once. This issue will be remedied by Hi-Art software version 4.0.3, scheduled for release in late 2010. Customers with questions were provided with a contact in the letter at the TomoTherapy Customer Interaction Center. The firm can be contacted at 608-824-2800.
- Initiated
- 2010-06-23
- Root cause
- Software design
- Terminated
- 2011-12-17
- Distribution
- CA, CO, IL, LA, MI, MO, NY, ND, OH, OK, PA,VA, WI, BELGIUM CANADA, KOREA, MALAYSIA, SWITZERLAND, ITALY, GREAT BRITAIN, FRANCE.
- Recall number
- Z-2072-2010
- Recall status
- Terminated
- CDRH record ID
- 92555
- FDA classified
- 2010-07-22
- Recall event number
- 56056