Open original FDA record ↗Recall record
- FDA identifier
- Z-2072-2015
- Company in FDA record
- Biomet, Inc.
- FDA classification date
- Jul 14, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medium Bone Plate Forceps was assembled with the incorrect grip arm tip. Potential that the surgeon will be unable to provisionally clamp the plate to the patient's bone.
Source facts
- Action
- On 6/17/2015, URGENT MEDICAL DEVICE RECALL NOTICE notifications were sent to the affected distributors with instructions for locating and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
- Initiated
- 2015-06-17
- Root cause
- Vendor change control
- Terminated
- 2016-03-22
- Distribution
- Domestic: TX NJ, NC, KY, MA, SD, MO, AL, IN, GA, AR International: Netherlands, Japan, Costa Rica
- Recall number
- Z-2072-2015
- Recall status
- Terminated
- CDRH record ID
- 138179
- FDA classified
- 2015-07-14
- Recall event number
- 71549