Z-2072-2017: GIZA Vertebral Body Replacement

Recall record Z-2072-2017 for Eden Spine Europe SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2072-2017
Company in FDA record
Eden Spine Europe SA
FDA record creation date
May 12, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaints that implants have been disassembled by surgeons because of unscrewing completely the locking screw.

Source facts

Action
A correction letter to the users was sent on March 28, 2017 Stating the following: Eden Spine Europe has become aware of extremely rare incidents of disassembly of Giza devices, which is distributed by Stryker Spine in the USA. These incidents have occurred in limited situations when a surgeon has needed to unscrew the device in order to readjust the distraction. To mitigate this risk, a more precise description of the unlocking procedure in the Giza Surgical Technique was added to prevent surgeons from fully unscrewing the locking screw. POTENTIAL HAZARD In the unlikely event the Giza device was to disassemble, the potential hazard is that the surgeon may decide either to reassemble the device or change the implant to a new device. This could potentially lead to a short surgical time delay of 2 to 10 minutes. RISK MITIGATIONS Eden Spine Europe and Stryker have added the following information in the Giza Surgical Technique to mitigate the risk of disassembly of the device: Note: If adjustment is needed, the implant locking mechanism must be unscrewed 3.5 to 4 rotations from the tightened position, to free the superior cylinder from the inferior one. No more than 4 counterclockwise rotations should be applied to the locking screw, as doing so could potentially lead to disassembly of the implant's locking mechanism. The GIZATM Vertebral Body Replacement surgical technique is available on Stryke Spine's website. ACKNOWLEDGEMENT OF THIS NOTICE Please acknowledge within 5 days of receipt of this notice: - Fill out and send the following online for: https://fr.surveymonkey.com/r/5XYT3ZQ - Or Send an Email to the following address: Customer.service@edenspine.com stating that this notice was received. For any questions you may contact: Christophe Zuanon, Regulatory Affairs & Quality Assurance Manager c.zuanon@edenspine.com Manager, c.zuanon@edenspine.com
Initiated
2017-03-17
Root cause
Labeling design
Terminated
2020-01-06
Distribution
AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WY
Recall number
Z-2072-2017
Recall status
Terminated
CDRH record ID
154405
FDA record created
2017-05-12
Recall event number
76852

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗