Open original FDA record ↗Recall record
- FDA identifier
- Z-2073-2010
- Company in FDA record
- Sorin Group USA, Inc.
- FDA classification date
- Jul 22, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Although accessories were designed for use with two models, the HLM C5 and the HLM S5, the submission that included references to the Mast Roller Pump, B-Care5, and the Electrical Venous Line Occluder accessories covered the new HLM S5 system. The three accessories were not cleared for use with the HLM C5.
Source facts
- Action
- The two customers who had received the suspect device accessories were identified and were contacted by letter on 6/18/2010. They were told to immediately discontinue use of the accessories and arrange for return of the products. They were also told that they could continue to use the CD5 Machine without the recalled accessories. A Product Recall Response Form was to completed by each consignee. Questions are directed to the firm at 800-221-7943.
- Initiated
- 2010-06-18
- Root cause
- PMA
- Terminated
- 2010-07-21
- Distribution
- CA, PA, TN.
- Recall number
- Z-2073-2010
- Recall status
- Terminated
- CDRH record ID
- 92727
- FDA classified
- 2010-07-22
- Recall event number
- 56139