Open original FDA record ↗Recall record
- FDA identifier
- Z-2073-2012
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jul 24, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Potential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments. In a very rare scenario, the RT Therapist will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.
Source facts
- Action
- The firm issued "Urgent: Medical Device Correction" letters delivered by the Siemens Customer Service Group or by certified mail to all its direct consignees, informing them of the affected product and providing instructions on the recall. The firm stated that release of this UI TH018/12/S to distribute the notifications began on June 25, 2012.
- Initiated
- 2012-06-25
- Root cause
- Software design
- Terminated
- 2012-10-18
- Distribution
- Worldwide Distribution, including the state of Florida (USA) and the countries of Germany, India, Poland, Thailand, and the Republic of Korea.
- Recall number
- Z-2073-2012
- Recall status
- Terminated
- CDRH record ID
- 110823
- FDA classified
- 2012-07-24
- Recall event number
- 62519