Skip to content

Public FDA record

Z-2073-2025: VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-cham

Recall record Z-2073-2025 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2073-2025
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Jul 2, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to run the OP-LO protocol leading to a potential delay in patient results.

Source facts

Action
Quidel Ortho notified consignees on about 06/12/2025 via FedEx letter. Distributors were instructed to discontinue distribution and discard inventory of any affected lots, send the provided customer letter and Confirmation of Receipt to any affected customers, and complete and return their Confirmation of Receipt. Customers were instructed that if their laboratory is unable to generate QC results within acceptable range using lots from GEN 15 of VITROS OP Reagent, discontinue using, render unusable, and discard your inventory of VITROS OP Reagent, GEN 15. Additionally, there were requested to complete and return the Confirmation of Receipt, maintain the notification with all records, and provide the notification if affected units were further distributed.
Initiated
2025-06-12
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Recall number
Z-2073-2025
Recall status
Open, Classified
CDRH record ID
214298
FDA classified
2025-07-02
Recall event number
97034

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.