Open original FDA record ↗Recall record
- FDA identifier
- Z-2074-2020
- Company in FDA record
- Dako Denmark A/S
- FDA classification date
- May 21, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe tray, the updated Basic User Guide will reduce the potential risk of false negative results in the case of leakage of buffer and/or reagents that may impact staining.
Source facts
- Action
- The firm issued letters dated 10/30/2019 via email or standard mail, or phone call beginning 11/5/2019 notifying the customer of the issue, providing the risk to health, and actions that needed to be taken. On 09/03/2020, Agilent Dako sent Urgent Field Safety Notices to customers who were asked to take the following actions: 1) Carefully read and communicate to those within your organization of the updates stated in this notification letter and attached Autostainer Link Basic User Guide (rev. 06). Your local sales team will confirm with you that you have read and acknowledged this notification. 2) Your Agilent representative should be contacted immediately for assistance in the event of leakage. 3) Please discard previous versions of the Autostainer Link Basic User Guide that you are currently using.
- Initiated
- 2019-11-05
- Root cause
- Nonconforming Material/Component
- Terminated
- 2022-05-17
- Distribution
- Distribution was nationwide. There was government/military distribution. Foreign distribution was made to Canada, Australia, Bangaladesh, Belarus, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Macedonia, Morocco, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Poland, Qatar, Romania, Russia, Saudi Arabia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Uruguay, and Vietnam.
- Recall number
- Z-2074-2020
- Recall status
- Terminated
- CDRH record ID
- 178414
- FDA classified
- 2020-05-21
- Recall event number
- 84465