Z-2074-2020: Dako Autostainer instruments: Autostainer Link 48, Model AS480; Autostainer Plus Link Instrument, Model

Recall record Z-2074-2020 for Dako Denmark A/S. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2074-2020
Company in FDA record
Dako Denmark A/S
FDA classification date
May 21, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Autostainer instrument may not have a syringe tray installed which could result in a small volume of excess buffer being applied to several slides. In addition to the syringe tray, the updated Basic User Guide will reduce the potential risk of false negative results in the case of leakage of buffer and/or reagents that may impact staining.

Source facts

Action
The firm issued letters dated 10/30/2019 via email or standard mail, or phone call beginning 11/5/2019 notifying the customer of the issue, providing the risk to health, and actions that needed to be taken. On 09/03/2020, Agilent Dako sent Urgent Field Safety Notices to customers who were asked to take the following actions: 1) Carefully read and communicate to those within your organization of the updates stated in this notification letter and attached Autostainer Link Basic User Guide (rev. 06). Your local sales team will confirm with you that you have read and acknowledged this notification. 2) Your Agilent representative should be contacted immediately for assistance in the event of leakage. 3) Please discard previous versions of the Autostainer Link Basic User Guide that you are currently using.
Initiated
2019-11-05
Root cause
Nonconforming Material/Component
Terminated
2022-05-17
Distribution
Distribution was nationwide. There was government/military distribution. Foreign distribution was made to Canada, Australia, Bangaladesh, Belarus, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Macedonia, Morocco, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Poland, Qatar, Romania, Russia, Saudi Arabia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Uruguay, and Vietnam.
Recall number
Z-2074-2020
Recall status
Terminated
CDRH record ID
178414
FDA classified
2020-05-21
Recall event number
84465

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗