Z-2075-2017: Navitrack System - OS Knee Universal, Orthopedic Stereotaxic Instrument

Recall record Z-2075-2017 for Orthosoft, Inc. dba Zimmer CAS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2075-2017
Company in FDA record
Orthosoft, Inc. dba Zimmer CAS
FDA classification date
May 12, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

ZIMMER CAS voluntarily conducted a retrospective recall of the Navitrack System - OS Knee Universal, due to software anomalies which were corrected with a new revision in version 2.2.1.3.

Source facts

Action
This is a retrospective report of a correction taken on 20 June 2008. The two software issues were corrected in the new software version 2.2.1.3. There were 164 copies of the affected software, 5 of which were only temporary licenses. There were 152 of the 159 software copies that had a full license that were confirmed to have installed the updated instructions per the correction notice for a total percentage of 96%. All sales representatives were made aware of the software issues by written notice. Both software issues were corrected in a new revision of the software with no remaining effect. There have been no complaints since the correction was implemented. Recall notices were e-mailed to customers dated 06/20/2008.
Initiated
2008-06-20
Root cause
Software design
Terminated
2017-05-15
Distribution
Domestic: CA, FL, GA, IL, IN, KS, MD, MN, MS, NC, NJ, NY, OR, PA, RI, SD, TN, TX, WA, WI Foreign: Germany, Spain, Czech Republic, Netherlands, Switzerland, New Zealand, Morocco, Brazil, Australia, Austria, France, Italy, Singapore, United Kingdom
Recall number
Z-2075-2017
Recall status
Terminated
CDRH record ID
154948
FDA classified
2017-05-12
Recall event number
75612

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗