Skip to content

Public FDA record

Z-2076-2019: U3-3 Solar Blue, Laptop integrated power supply used with the Solar Blue Urodynamics for assessing lower

Recall record Z-2076-2019 for LABORIE MEDICAL TECHNOLOGIES, CORP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2076-2019
Company in FDA record
LABORIE MEDICAL TECHNOLOGIES, CORP
FDA record creation date
Jul 25, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire hazard

Source facts

Action
Laborie issued letter via FedX on July 17, 2019 advising of the problem, health risk, and action to take: Quarantine the product and unplug the power supply from the wall socket until a replacement is provided. A new power supply unit(s) will be provided along with instructions on how to replace the power supplie in the field. Contact Laborie/MMS customer care with any questions, concerns or problems: Laborie MMS Customer Care Address: 400 Avenue D, Suite 10, Williston, VT 05495 Phone Number: 800.522.6743 Email: customercareusa@laborie.com
Initiated
2019-07-17
Root cause
Nonconforming Material/Component
Terminated
2023-07-10
Distribution
US distribution in AZ, and PA.
Recall number
Z-2076-2019
Recall status
Terminated
CDRH record ID
174669
FDA record created
2019-07-25
Recall event number
83310

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗