Open original FDA record ↗Recall record
- FDA identifier
- Z-2076-2019
- Company in FDA record
- LABORIE MEDICAL TECHNOLOGIES, CORP
- FDA record creation date
- Jul 25, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire hazard
Source facts
- Action
- Laborie issued letter via FedX on July 17, 2019 advising of the problem, health risk, and action to take: Quarantine the product and unplug the power supply from the wall socket until a replacement is provided. A new power supply unit(s) will be provided along with instructions on how to replace the power supplie in the field. Contact Laborie/MMS customer care with any questions, concerns or problems: Laborie MMS Customer Care Address: 400 Avenue D, Suite 10, Williston, VT 05495 Phone Number: 800.522.6743 Email: customercareusa@laborie.com
- Initiated
- 2019-07-17
- Root cause
- Nonconforming Material/Component
- Terminated
- 2023-07-10
- Distribution
- US distribution in AZ, and PA.
- Recall number
- Z-2076-2019
- Recall status
- Terminated
- CDRH record ID
- 174669
- FDA record created
- 2019-07-25
- Recall event number
- 83310