Open original FDA record ↗Recall record
- FDA identifier
- Z-2078-2012
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jul 26, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens became aware of a potential malfunction when using syngo.plaza with software version VA20C_HF01. For datasets with distance measurements it can happen, that all images are not loaded to Viewer but no error message is displayed to user.
Source facts
- Action
- Siemens sent a Customer Safety Advisory Notice letter dated February 24, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to ensure that the prefix defined in the viewer settings for "Content Creator's Initials" is not empty. Then all images will be display correctly. This has to be checked for every user. Further questions please call 610-219-6300.
- Initiated
- 2012-02-24
- Root cause
- Software design
- Terminated
- 2014-01-14
- Distribution
- (USA) nationwide including the states of CA, FL, GA, MA, MN, MO, NJ, NY, OH, WA, and WI.
- Recall number
- Z-2078-2012
- Recall status
- Terminated
- CDRH record ID
- 110712
- FDA classified
- 2012-07-26
- Recall event number
- 62462