Z-2078-2025: REF: 335775 , BD FACS, 7-Color Setup Beads, 25 setups per kit, Rx Only, IVD. Used in flow cytometry for f

Recall record Z-2078-2025 for Becton, Dickinson and Company, BD Bio Sciences. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2078-2025
Company in FDA record
Becton, Dickinson and Company, BD Bio Sciences
FDA classification date
Jul 3, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.

Source facts

Action
On June 3, 2025, Becton, Dickinson issued a "Urgent: Medical Device Recall Correction" Notification to affected customers via mail and E-Mail. BD asked consignees to take the following action: 1: Immediately inspect your inventory for the specific catalog and lot numbers listed. It is recommended that the clinical laboratory staff review and segregate the affected lots. 2. Remove and discard prior affected label insert with incorrect spectral overlap factor values included on kit box prior to use. 3. The wrong SOF (Spectral Over Factors) values can be corrected by entering the right SOF values provided in Table 1. 4. Ensure the right SOF values are entered in the flow cytometry instrument prior to sample acquisition. 5. Customers who have not used the affected product may proceed to enter the SOF values. 6. If the affected products have been fully or partially used and subjects' samples have been processed and acquired on an instrument with the incorrect SOF values, please review the results. 7. If the subject sample(s) is not available anymore, this may require the subject to return to collect an additional sample for testing purposes. 8. Ensure the contents of this Product Recall Notification are read and understood. 9. Share and post this customer letter with the users of this application at your facility and forward to any customers to whom you may have distributed the product to ensure awareness.
Initiated
2025-06-03
Root cause
Process control
Distribution
US: Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Dakota, Tennessee, Texas, Utah, Virginia, Wisconsin, West Virginia Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, Hong Kong, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, Peru, Pakistan, Singapore, El Salvador, Taiwan OUS:
Recall number
Z-2078-2025
Recall status
Open, Classified
CDRH record ID
214040
FDA classified
2025-07-03
Recall event number
96922

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗