Open original FDA record ↗Recall record
- FDA identifier
- Z-2081-2008
- Company in FDA record
- Varian Medical Systems, Inc.
- FDA classification date
- Sep 16, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Medical device may exhibit inaccurate positioning and lead to unintended dose delivery during brachytherapy treatment.
Source facts
- Action
- The firm notified medical device users by letter on 03/12/08. The notification provided updated instructions for use for the CT/MR Ring and Tandem Combination Applicator Set. Users were advised to characterize all ring applicators to develop an offset correction for each dwell position in order to construct a plan to deliver the dwell positions accurately and reliably. The characterization can be done by creating a set of autoradiographs with intended dwell positions and define treatment plan shifts in order to deliver dwell positions in intended locations. Users were advised that a subsequent set of verification autoradiographs should be made using the derived plan shifts to verify the corrected source positioning. The notification also advised that an individual ring's behavior can change over its duration of use and that periodic commissioning is dictated to ensure reproducibility and compensate for changes in the device over time. Users were instructed to inform all employees and extend the notice to any consignees. Users were requested complete and return the enclosed response form. If you have questions about this recall, send an email to brachyhelp@varian,com.
- Initiated
- 2008-03-12
- Root cause
- Device Design
- Terminated
- 2012-11-28
- Distribution
- The medical devices were distributed to radiation oncology facilities nationwide.
- Recall number
- Z-2081-2008
- Recall status
- Terminated
- CDRH record ID
- 71103
- FDA classified
- 2008-09-16
- Recall event number
- 48219