The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed.
Source facts
Action
Alcon has already notified the affected customer verbally on May 24, 2016, and via written correspondence on June 1, 2016. The device was corrected on May 26, 2016, and the Power Entry Switch that was removed from the customers CENTURION Vision System did not exhibit a grounding pin susceptible to detachment. For further questions, please call (817) 293-0450.
Initiated
2016-05-24
Root cause
Nonconforming Material/Component
Terminated
2017-04-13
Distribution
US Distribution to Illinois
Recall number
Z-2081-2016
Recall status
Terminated
CDRH record ID
147055
FDA classified
2016-06-28
Recall event number
74282