One bolt on the V motor was not fully tightened which caused a drift of .13 degrees in alignment resulting in reduced dose output.
Source facts
Action
The firm sent a Field Correction Notice letters dated 7/3/19 to customers. The letter identified the affected product, problem and actions to be taken. If you need any further information or support concerning this issue, please contact Xcsion Customer Support at 855-924-7466.
Initiated
2019-07-03
Root cause
Nonconforming Material/Component
Terminated
2021-07-01
Distribution
The products were distributed to the following US states: MD and TX.
Recall number
Z-2081-2019
Recall status
Terminated
CDRH record ID
174834
FDA record created
2019-07-25
Recall event number
83342