Z-2082-2013: 2008 MeDS Pump, Venofer(R) Pump
Product Usage: The 2008 MeDs pump is an optional accessory for use with
Recall record Z-2082-2013 for Fresenius Medical Care Holdings, Inc. dba Renal Solutions. View source facts, labeled dates, and the original FDA record.
Fresenius Medical Care Holdings, Inc. dba Renal Solutions
FDA classification date
Aug 26, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
Product is being recalled because the combination of the current 50 mg vial, the 2008 MeDS pump, and the 2008 MeDS pump tubing Set for CombiSet Bloodlines can lead to a situation where Venofer(R) delivery does not meet dosage accuracy requirements. This may lead to a slight over or under delivery of Venofer(R) that does not pose an acceptable risk to the patient.
Please note that this recall
Source facts
Action
Renal Solutions sent an Urgent Notification letter dated April 1, 2013 via certified mail to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to cease further use of the device. The letter informed customers that a Fresenius Medical Care Technical Service/Customer Service representative will contact customers to arrange for removal of the device. For questions contact Fresenius Medical Care Customer Service at 800-227-2572.
Initiated
2013-04-01
Root cause
Device Design
Terminated
2015-03-02
Distribution
USA Nationwide Distribution in the states of CA, MA, and NY.
Recall number
Z-2082-2013
Recall status
Terminated
CDRH record ID
117850
FDA classified
2013-08-26
Recall event number
65080