Z-2082-2017: HDL Cholesterol Reagent (HDL 2 x 200), Catalog No. A15625. Intended for quantitative determination of HD

Recall record Z-2082-2017 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2082-2017
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
May 12, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.

Source facts

Action
An Urgent Medical Device Recall letter dated 4/3/17 was sent to customers to inform them that Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole, Sulfasalazine, and Sulfapyridine. Customers are informed that with the administration of these drugs, the effects to patient analytes are: - Metamizole (Dipyrone) may generate false low Cholesterol, Enzymatic Creatinine, HDL Cholesterol, Lactate, Triglycerides and Uric Acid results. -Sulfasalazine may generate false high Salicylate results. -Sulfapyridine may generate false low Salicylate results. Customers are informed of the action and resolution. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with any questions regarding the notice are instructed to contact Customer Support Center -Via http://www.beckmancoulter.com/customersupport/support -Via phone, call 1-800-854-3633 in the United States and Canada Customers outside the US and Canada are instructed to contact their local Beckman Coulter Representative.
Initiated
2017-04-03
Root cause
Under Investigation by firm
Terminated
2018-05-31
Distribution
Worldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, United States including Puerto Rico, and Vietnam.
Recall number
Z-2082-2017
Recall status
Terminated
CDRH record ID
154701
FDA record created
2017-05-12
Recall event number
76962

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗