Z-2083-2016: Medtronic Navigation O-arm 02 surgical x-ray imaging system

Recall record Z-2083-2016 for Medtronic Navigation, Inc.-Littleton. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2083-2016
Company in FDA record
Medtronic Navigation, Inc.-Littleton
FDA record creation date
Jul 6, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Review of on-system electronic product labeling and Instr uctions for Use in response to an internal management concern uncovered gaps in compliance with the requirements for information to be provided to users in 21CFR part 1 020.30(h).

Source facts

Action
Medtronic planned action to bring device into Compliance: 1. You will contact customers and provide an Errata sheet to them updating the missing and incorrect information. 2. Medtronic field service personnel will perform radiation output measurements via validated procedures to attain compliance with 21CRF part 1020.32(k)(6). 3. You will update the systems software-based instructions for use by a software update which includes the information supplied by the Errata sheet. For questions call (800)-595-9709.
Initiated
2016-03-17
Root cause
Radiation Control for Health and Safety Act
Terminated
2021-04-06
Distribution
Nationwide Distribution
Recall number
Z-2083-2016
Recall status
Terminated
CDRH record ID
146536
FDA record created
2016-07-06
Recall event number
74158

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗