Open original FDA record ↗Recall record
- FDA identifier
- Z-2083-2016
- Company in FDA record
- Medtronic Navigation, Inc.-Littleton
- FDA record creation date
- Jul 6, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Review of on-system electronic product labeling and Instr uctions for Use in response to an internal management concern uncovered gaps in compliance with the requirements for information to be provided to users in 21CFR part 1 020.30(h).
Source facts
- Action
- Medtronic planned action to bring device into Compliance: 1. You will contact customers and provide an Errata sheet to them updating the missing and incorrect information. 2. Medtronic field service personnel will perform radiation output measurements via validated procedures to attain compliance with 21CRF part 1020.32(k)(6). 3. You will update the systems software-based instructions for use by a software update which includes the information supplied by the Errata sheet. For questions call (800)-595-9709.
- Initiated
- 2016-03-17
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2021-04-06
- Distribution
- Nationwide Distribution
- Recall number
- Z-2083-2016
- Recall status
- Terminated
- CDRH record ID
- 146536
- FDA record created
- 2016-07-06
- Recall event number
- 74158