Open original FDA record ↗Recall record
- FDA identifier
- Z-2083-2019
- Company in FDA record
- Randox Laboratories, Limited
- FDA record creation date
- Jul 25, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.
Source facts
- Action
- On June 24, 2019, the firm notified its customers of the recall through an Urgent Field Safety Notice. Customers were asked to take the following actions: If there is any sign of damage to the RCU lid or lid sensor " Switch of main analyser main power supply as soon as possible " Contact technical support immediately. " Discuss the contents of this notice with your Medical Director. " Display this notice in a prominent location near the instrument. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days.
- Initiated
- 2019-06-24
- Root cause
- Under Investigation by firm
- Terminated
- 2020-05-21
- Distribution
- Domestic distribution to CA CT MA MD MO NC NJ NY OH PA UT WV and PR. Worldwide foreign distribution.
- Recall number
- Z-2083-2019
- Recall status
- Terminated
- CDRH record ID
- 173928
- FDA record created
- 2019-07-25
- Recall event number
- 83175