Open original FDA record ↗Recall record
- FDA identifier
- Z-2084-2012
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- Jul 26, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Stryker Orthopaedics received a report indicating that a Tibial Keel Punch could not be inserted into the slot of the Tibial Punch Tower. The slot that guides the stem of the keel punch into the channel of the tower was later determined to be undersized.
Source facts
- Action
- Stryker Orthopaedics sent an Urgent Product Recall/Recall Acknowledgement letter dated January 10, 2012, to all affected customers. The letter identified the product(s) the problem and the action needed to be taken by the customer. Customers were instructed to contact their Stryker Sales Representative to arrange for return of the affected product. Customers were asked to fax back the attached Product Recall Acknowledgement Form as soon as possible to 201-831-6069. For questions customers should call 201-972-2100. For questions regarding this recall call 201-831-5970.
- Initiated
- 2011-12-22
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-06-13
- Distribution
- Worldwide Distribution
- Recall number
- Z-2084-2012
- Recall status
- Terminated
- CDRH record ID
- 110358
- FDA classified
- 2012-07-26
- Recall event number
- 62365