Skip to content

Public FDA record

Z-2084-2014: RECLAIM DISTAL TAPERED Distal Stem, various sizes. Taper Protector Sleeve. RECLAIM Revision Hip System O

Recall record Z-2084-2014 for DePuy Orthopaedics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2084-2014
Company in FDA record
DePuy Orthopaedics, Inc.
FDA classification date
Jul 21, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The product can be difficult to remove from the Distal Stem both out of the package and after proximal reaming.

Source facts

Action
DePuy Synthes sent an Urgent Information - Device Correction Notice letter dated June 24, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: 1. Immediate Remedy: Discard the RECLAIM Taper Sleeve Protector instrument before stem implantation. 2. Changes to the literature and IFU: Documents are revised as follows: " RECLAIM Revision Hip Systems Surgical Technique # DSUS/JRC/0614/0200 pages 11, 12, 13, 16, 30, 34, 35 (figure numbering from pages 11 to 27 were also revised): Removed information regarding taper sleeve protector. " RECLAIM Revision Hip Systems Surgical Technique FAQ # DSUS/JRC/0614/0202 pages 1 and 2: Removed information regarding taper sleeve protector. " RECLAIM Revision Hip Systems IFU #090200799 REV E Page 5: Verbiage will be revised and implemented as soon as possible, depending on regulatory approvals. 3. Change: The company consulted with design surgeons and developed a strategy to eliminate the taper sleeve protector from the system. As a result of removing the taper sleeve protector, the company is evaluating whether there is a need for a protective component to be used during the distribution Customers with questions were instructed to call 574-371-4917. For questions regarding this recall call 574-371-4577.
Initiated
2014-06-03
Root cause
Device Design
Terminated
2017-05-17
Distribution
Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI & WV and Internationally to AUSTRALIA, AUSTRIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, IBERIA, IRELAND, ISRAEL, ITALY, NEW ZEALAND, NORWAY, POLAND, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWITZERLAND & UK.
Recall number
Z-2084-2014
Recall status
Terminated
CDRH record ID
128022
FDA classified
2014-07-21
Recall event number
68440

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗