Open original FDA record ↗Recall record
- FDA identifier
- Z-2084-2016
- Company in FDA record
- Integra LifeSciences Corp.
- FDA record creation date
- Jun 28, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stainless steel instead of 304 stainless steel. b) K-wires have passivated surface finish instead of electro-polish surface finish. c) K-wires have incorrect length and diameter.
Source facts
- Action
- Integra sent an Urgent Voluntary Medical Device Recall letter dated May 20, 2016, to all affected consignees regarding the recall via traceable courier service. The letter identified the product, the problem, and the action to be taken by the customer. Customers who have supplied affected product should forward a copy of the Recall Notice to them. Customers should review their inventory for affected product, complete the form as to whether they have any affected product or not, return the form by email or fax as indicated on the form. Customers with questions should contact Customer Service at 1-800-654-2873. For questions regarding this recall call 609-936-6822.
- Initiated
- 2016-05-20
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-11-16
- Distribution
- Nationwide Distribution to OK, CA, TX, WA, GA, NY, MS, NY, PA
- Recall number
- Z-2084-2016
- Recall status
- Terminated
- CDRH record ID
- 146870
- FDA record created
- 2016-06-28
- Recall event number
- 74224