Skip to content

Public FDA record

Z-2084-2019: RX Imola, Model Nos. RX4900

Recall record Z-2084-2019 for Randox Laboratories, Limited. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2084-2019
Company in FDA record
Randox Laboratories, Limited
FDA record creation date
Jul 25, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.

Source facts

Action
On June 24, 2019, the firm notified its customers of the recall through an Urgent Field Safety Notice. Customers were asked to take the following actions: If there is any sign of damage to the RCU lid or lid sensor " Switch of main analyser main power supply as soon as possible " Contact technical support immediately. " Discuss the contents of this notice with your Medical Director. " Display this notice in a prominent location near the instrument. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days.
Initiated
2019-06-24
Root cause
Under Investigation by firm
Terminated
2020-05-21
Distribution
Domestic distribution to CA CT MA MD MO NC NJ NY OH PA UT WV and PR. Worldwide foreign distribution.
Recall number
Z-2084-2019
Recall status
Terminated
CDRH record ID
173929
FDA record created
2019-07-25
Recall event number
83175

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗