Z-2086-2014: NOVATION Femoral Stem, Tapered Extended offset, plasma- coated, press fit. Catalog number 160-01-13 in

Recall record Z-2086-2014 for Exactech, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2086-2014
Company in FDA record
Exactech, Inc.
FDA classification date
Jul 21, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The femoral stem is lacking specified plasma coating.

Source facts

Action
Exactech sent an Important Product Recall Notice dated March 10, 2014, to all affected customers. The notice identified the product the problem and the action to be taken by the customer. Description of Issue: The femoral stem is lacking specified plasma spray coating. Clinical Impact: Without plasma spray coating, the stem may not remain stable and will not have the possibility for biological fixation. In order to comply with applicable regulations and Exactech policies customers were instructed to : " Immediately cease distribution or use of these products. " Extend this information to your accounts that may have this product in their possession. " Determine if they have any of the affected devices " Fax back the attached form. Complete the attached fax back form and return it to Exactech. Customers were instructed to contact their Exactech inventory representative within 5 business days to confirm quantities at their location. For any inventory restocking questions related to this issue, please call 1-800-392-2832. For questions regarding this recall call 800-392-2832.
Initiated
2014-03-07
Root cause
Component design/selection
Terminated
2014-08-04
Distribution
Worldwide Distribution including Spain, Greece and Japan
Recall number
Z-2086-2014
Recall status
Terminated
CDRH record ID
127375
FDA classified
2014-07-21
Recall event number
68341

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗