Z-2087-2019: Maximum Pelvic Trainer (MPT)

Recall record Z-2087-2019 for Ralston Group. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2087-2019
Company in FDA record
Ralston Group
FDA record creation date
Jul 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The device was incorrectly listed with the FDA and is therefore considered adulterated and misbranded if left on the market. Additionally, the firm did not submit a 510k for the device.

Source facts

Action
Ralston Group notified customers on about 07/12/2019 via "IMPORTANT VOLUNTARY: MEDICAL DEVICE RECALL" email. Customers were advised that the device was incorrectly registered with the FDA and is therefore considered adulterated and misbranded if left on the market. Processes and documentation regarding the manufacture of this device were not adequately maintained. Also, clinical trials and claims concerning the effectiveness of the MPT were not evaluated by the FDA. Instructions included to immediately examine inventory for affected product, discontinue use/distribution, and quarantine al affected product. Customers were requested to complete the provided MEDICAL DEVICE RECALL RETURN RESPONSE to confirm receipt and understanding, and arrange for return of all affected product by contacting customer service at info@bestforlife.com or call 1-800-867-3726, Monday -Friday 8 AM 5 PM Central Time. Also customers were instructed to notify customers if the product was further distributed.
Initiated
2019-07-12
Root cause
Other
Terminated
2022-04-15
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV and PR.
Recall number
Z-2087-2019
Recall status
Terminated
CDRH record ID
174733
FDA record created
2019-07-26
Recall event number
83328

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗