Skip to content

Public FDA record

Z-2088-2019: Filter-Tips, 1500 ul (1024 tips/kitbox), REF 997024- REF: 997024 IVD: Used with the QIAGEN QIAsymphony S

Recall record Z-2088-2019 for Qiagen Sciences LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2088-2019
Company in FDA record
Qiagen Sciences LLC
FDA classification date
Aug 1, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Kit lots contain a small percentage of damaged Filter-Tips that are leaking when being used with the QIAsymphony SP/AS Instruments (REF 9001297 and 9001301)

Source facts

Action
QIAGEN issued Urgent Device Notification dated June 5, 2019 stating problem, health risk and action to take: Forward this information to all individuals and departments who are using the above listed kit within your organization. If you are not the end user, please forward this notice to the product end user. If you have any remaining stocks of Filter-Tips, 1500 l (1024) REF 997024, lots 0605020019 and 0605020020, DO NOT USE them. For a free-of-charge replacement, please contact QIAGEN Technical Service. Dispose of the product in accordance with your national and local safety and environmental regulations. l Review this notice with your laboratory/medical director. l If you have used the affected kits with lots 0605020019 and/or 0605020020, we recommend a review of the corresponding results. l Complete the Acknowledgement of Receipt form attached to this letter by June 12, 2019 and email it to quality.communications@qiagen.com. Questions or concerns, contact your local QIAGEN Technical Service Department through any of the following: Telephone: 800 362 7737 Email: TechService-NA@qiagen.com
Initiated
2019-06-05
Root cause
Process control
Terminated
2020-08-28
Distribution
US Nationwide distribution
Recall number
Z-2088-2019
Recall status
Terminated
CDRH record ID
173998
FDA classified
2019-08-01
Recall event number
83199

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗