Open original FDA record ↗Recall record
- FDA identifier
- Z-2088-2025
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Jul 3, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
Source facts
- Action
- An Urgent Medical Device Correction notification dated 6/4/25 was mailed to consignees. This notification provides consignees with calibration instructions given the issue found in affected lots. Additionally, instructions are given on how to proceed with product depending on performance. Consignees are to discontinue use of and discard of the calibrator SETPOINT lots impacted by this recall. Consignee locations are to review the provided notification with their Medical Directors to determine the appropriate course of action. A copy of the recall notification should be retained in laboratory records and shared with users. Consignees are to complete and return the provided Effectiveness Check Form within 30 days of receipt of the recall notification. Consignees with any questions can contact their Siemens Customer Care Center or local Siemens Healthineers technical support representative.
- Initiated
- 2025-06-04
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution - US Nationwide and the countries of AR, AU, BR, CA, CL, CN, CO, CR, HK, ID, IN, JP, KR, MX, MY, PE, PY, SA, SG, TH, TW, UY, VN, & ZA.
- Recall number
- Z-2088-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214306
- FDA classified
- 2025-07-03
- Recall event number
- 97037