The firm issued a clinical alert after users reported an issue where the Anesthesia-Rx system was interrupted ANF became temporarily unavailable during use.
Source facts
Action
McKesson sent a Clinical Alert letter dated July 22, 2013 to all affected customers. The letter identified the affected product, symptoms, and actions to be taken. McKesson committed to releasing 8.0.2 Update 1.5 to affected customers which will include enhanced code to prevent the unhandled excpetion errors. Additional resolution steps were included in the letter. For questions call 1-800-700-8737.
Initiated
2013-07-22
Root cause
Software design
Terminated
2014-06-19
Distribution
USA Nationwide Distribution in the states AR, MD, NJ, NY, PA, and WV.
Recall number
Z-2089-2013
Recall status
Terminated
CDRH record ID
120087
FDA classified
2013-08-26
Recall event number
65769