Z-2090-2019: Flexible Drill Shaft, FMP Acetabular System Instrumentation REF 803-05-239 RX Encore Medical, L.P.

Recall record Z-2090-2019 for Encore Medical, Lp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2090-2019
Company in FDA record
Encore Medical, Lp
FDA record creation date
Jul 27, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Manufacturing error, the drill bit shaft is unable to properly capture the drill bit pin. Failure of the drill bit shaft to capture the drill bit pin could result in the drill bit coming loose which could result in a delay in surgery or adverse patient reaction resulting in a revision surgery.

Source facts

Action
On 9/19/2018 the firm emailed an "Urgent Field Safety Notice" letter to their consignees. The Field Safety Notice letter asked the customers to do the following: 1.Contact their customer service to request a RMA for return and exchange. 2.If the customer does not have the drill shaft or the number of sets identified, please contact Joan Busick at andra.busick@djdoglobal.com. 3. If the customer has questions they should contact Teffany Hutto by phone at (512) 834-6255.
Initiated
2018-09-19
Root cause
Nonconforming Material/Component
Distribution
US ( MS, NY, MO, OK, UT, AZ)
Recall number
Z-2090-2019
Recall status
Open, Classified
CDRH record ID
173392
FDA record created
2019-07-27
Recall event number
83037

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗