Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.
Source facts
Action
The firm initiated the recall to the consignee level on January 15, 2024. Customers were asked to coordinate a cartridge replacement. The reporting of this recall was done by the firm retroactively; the issue was resolved in January 2024.
Initiated
2024-01-19
Root cause
Software design
Distribution
US Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Maryland, Massachusetts,
Michigan, Nebraska, New York, North Carolina, North Dakota, Oklahoma, Oregon, Utah,
Wisconsin.
Recall number
Z-2090-2025
Recall status
Completed
CDRH record ID
214308
FDA classified
2025-07-07
Recall event number
97021