Skip to content

Public FDA record

Z-2090-2025: TrueDent Clear, TDM 100, Model No. OBJ-09164 Light-curable methacrylate-based resin that enables fabricat

Recall record Z-2090-2025 for STRATASYS LTD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2090-2025
Company in FDA record
STRATASYS LTD
FDA classification date
Jul 7, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.

Source facts

Action
The firm initiated the recall to the consignee level on January 15, 2024. Customers were asked to coordinate a cartridge replacement. The reporting of this recall was done by the firm retroactively; the issue was resolved in January 2024.
Initiated
2024-01-19
Root cause
Software design
Distribution
US Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Maryland, Massachusetts, Michigan, Nebraska, New York, North Carolina, North Dakota, Oklahoma, Oregon, Utah, Wisconsin.
Recall number
Z-2090-2025
Recall status
Completed
CDRH record ID
214308
FDA classified
2025-07-07
Recall event number
97021

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗