Z-2091-2023: Aptio Automation Storage and Retrieval Module (SRM)

Recall record Z-2091-2023 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2091-2023
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Jul 5, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a delay in testing, may occur when all the conditions listed below occur within a few milliseconds timeframe of each other: - The SRM module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the SRM module buffer - There is divert gate malfunction and a narrow timing window Only in this specific scenario, Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without a unique error message being generated. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the pending test orders (not yet performed) on Tube B.

Source facts

Action
The firm sent a recall letter dated 05/16/2026 to consignees notifying them of the firmware issue presenting with this recall event. Consignees are asked to conduct visual checks of the device's gates at the buffer lane entry daily to ensure there are no obstructions that prevent the correct activation and movement of the divert, the divert looks intact, and the position of the divert is aligned to the profiles when it is diverting a tube without any missing or partial blocking of the carrier. If the visual check does not pass, contact your local technical support for assistance before using the impacted system. Until a service visit to upgrade the firmware of affected devices can be conducted, consignees are asked to maintain awareness of the provided recall notice and conduct daily recommended actions. Consignees are asked to complete and return the provided response form within 7 days of recall notification. Completed forms are to be emailed to escctsfcaecfax.team@siemens-healthineers.com or faxed to (312)274-7795. Customers with any questions are asked to contact their local Siemens Healthineers Customer Care Center or technical support representative.
Initiated
2023-05-16
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, CO, FL, IL, KY, LA, MA, MD, MI, MS, MT, NC, NH, NY, OH, PA, SC, TN, TX, VA, WA, & WV. The countries of Australia, Brazil, Canada, China, France, Germany, Greece, Italy, Portugal, Saudi Arabia, Spain, Sweden, & United Kingdom.
Recall number
Z-2091-2023
Recall status
Open, Classified
CDRH record ID
200970
FDA classified
2023-07-05
Recall event number
92461

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗