Open original FDA record ↗Recall record
- FDA identifier
- Z-2092-2014
- Company in FDA record
- Alphatec Spine, Inc.
- FDA classification date
- Jul 22, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Alphatec Spine is recalling the 2.1mm Fixed Depth Drill 10mm because they received a report that the instrument malfunctioned and golden stop slid from its fixed position during use.
Source facts
- Action
- Alphatec Spine, Inc. sent an Urgent Medical Device Notification letter dated June 20, 2014, June 26, 2014, to all affected customers. Alphatec Spine received a report of the 10mm fixed depth drill from this lot which malfunctioned and golden stop slid from its fixed position during use. The letter informs the customers of the problems identifed and the actions to be taken. Customers are instructed to confirm that they received the notification letter by completing the customer response form. Completed forms are to be sent to Alphatec Spine at 5818 El Camino Rd, Carlsbad, CA 92008. Customers may also return forms by fax at (760) 431-0289. Customers with questions were instructed to contact Alphatec Spine's Customer Service Department on how to return the affected product. For questions regarding this recall call 760-431-9286.
- Initiated
- 2014-06-26
- Root cause
- Unknown/Undetermined by firm
- Terminated
- 2014-09-03
- Distribution
- Nationwide Distribution including CT, NC, AL, WV, IL, AR, and NY.
- Recall number
- Z-2092-2014
- Recall status
- Terminated
- CDRH record ID
- 128421
- FDA classified
- 2014-07-22
- Recall event number
- 68667