Open original FDA record ↗Recall record
- FDA identifier
- Z-2092-2015
- Company in FDA record
- Synthes, Inc.
- FDA classification date
- Jul 16, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was reported that the RIA Drive Shaft, Tube Assembly, and Reamer Head have the potential to break when incorrectly assembled or used improperly. Precautionary statements are being added to the Reamer / Irrigator / Aspirator (RIA) Surgical Technique Guide.
Source facts
- Action
- DePuy Synthes sent an Urgent Medical Device Safety Notification dated May 28, 2015, to direct accounts and sales consultants to advise them of the patient risk associated with misassembly or misuse of the device. Customers and sales force were instructed to remove the Surgical Technique Guide J4352 (all versions) dated 11/2011 or earlier and replace it with the updated Technique Guide (DSUS/TRM/0515/0620, Date 06/2015),and to return the completed verification response form. If you have any further questions, please call (610) 719-5450. Thank you for your attention to this matter.
- Initiated
- 2015-05-28
- Root cause
- Labeling False and Misleading
- Terminated
- 2016-08-18
- Distribution
- Nationwide Distribution and Internationally to Canada and Switzerland.
- Recall number
- Z-2092-2015
- Recall status
- Terminated
- CDRH record ID
- 137455
- FDA classified
- 2015-07-16
- Recall event number
- 71366