Skip to content

Public FDA record

Z-2092-2021: Hemotherm CE Dual Reservoir Cooler/Heater, Model 400CE, 115V Model, part no. 86022. Used to cool or warm

Recall record Z-2092-2021 for Gentherm Medical, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2092-2021
Company in FDA record
Gentherm Medical, LLC
FDA record creation date
Jul 19, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential risk of device contamination and patient infection associated with the device because of the potential for organisms (including Nontuberculous mycobacteria NTM) to grow in the water systems of any heater-cooler device, and contaminated water from any heater-cooler device has the potential to aerosolize into the operating room during surgery which could lead to patient infection.

Source facts

Action
On 6/17/2021, Gentherm issued Urgent Medical Device Recall notices and response forms to customers via courier service. Instructions to Customers: 1) Immediately examine your inventory, access updated manuals and ensure obsolete manuals are removed from service. 2) Updated manuals may be accessed via www.gentherm.com or physicial copies may be requested from Gentherm Medical, LLC at 1-888-437-5608. 3) Ensure that all users are informed of the contents of this letter. 4) If you have further distributed this product, please provide those accounts with a copy of this notice. 5) Complete and return the customer response form via email to FA2021-004@gentherm.com.
Initiated
2021-06-17
Root cause
Process control
Terminated
2023-03-01
Distribution
Worldwide distribution. US nationwide, including Puerto Rico; Argentina, Bangladesh, Chile, Colombia, France, Guatemala, Hungary, India, Indonesia, Israel, Italy, Nepal, Nigeria, Pakistan, Philippines, Romania, Saudi Arabia, Singapore, South Africa, South Korea, Switzerland, Taiwan, Thailand, Turkey, UAE
Recall number
Z-2092-2021
Recall status
Terminated
CDRH record ID
188058
FDA record created
2021-07-19
Recall event number
88187

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗