Open original FDA record ↗Recall record
- FDA identifier
- Z-2093-2017
- Company in FDA record
- Del Mar Reynolds Medical, Ltd.
- FDA record creation date
- May 16, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.
Source facts
- Action
- Spacelabs sent an URGENT - MEDICAL DEVICE CORRECTION letter dated April 13, 2017, to US customers via priority service. The customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record on April 14, 2017. The letter announced that the The Arkon Anesthesia Delivery System, Model 99999, and the Absorber Adapter Assembly Kit part number 050-9045-00, and Absorber Adapter Assembly part number 650-1648-01 are being recalled. Customers are advised that The Arkon Anesthesia Delivery Systems Model 99999 with Absorber Adapter Assembly part number 650-1648-00 is NOT affected by this recall. Spacelabs Healthcare is developing a hardware upgrade to improve the robustness of the affected Absorber Adapter Assembly. A Spacelabs representative will contact consignees once the upgrade is available to schedule a convenient time to update their Arkon Anesthesia Delivery System(s) at no cost. For additional information or technical assistance, please contact: 1-800-522-7025 select 4 for Anesthesia Technical Support
- Initiated
- 2017-04-14
- Root cause
- Device Design
- Terminated
- 2017-10-16
- Distribution
- US distribution in FL, LA, MS, NC, NC, SC and WY and Internationally to Panama and United Kingdom.
- Recall number
- Z-2093-2017
- Recall status
- Terminated
- CDRH record ID
- 154830
- FDA record created
- 2017-05-16
- Recall event number
- 77019